Exchange Transfusion
|
Reviewed by K. Brokenshire & Pita Birch |
| December 2009 |
| Step | Action |
| 1 | Use prepacked Exchange Transfusions two catheter push-pull set. See diagram below. |
| 2 | Place Astroflow unit on to pole on ponta, switch unit on and adjust setting to 39.5° C, then press start (confirms the change or unit will reset to default settings) |
| 3 | Prime the blood filter, Alaris giving set and one way valve with blood. |
| 4 | When ready to commence the exchange, connect the outline to baby’s arterial line. |
| 5 | Place the Alaris line in the warming sheath (from the roller clamp down) and connect to the baby's venous line. |
| 6 | The nurse assisting may manage one of the lines as directed by the Consultant/Registrar/NS-ANP. (If this happens another level 4 nurse must then be doing the recording of blood volumes) |
| 7 | The nurse is responsible for recording in and out blood volumes. Each cycle in and out should take approximately 4 minutes. |
| N.B. |
|
Observe carefully throughout the procedure that there is no air in the line.
Click on the image above to open the image