Nitric Oxide
Administration of Nitric Oxide

 

Authorised by:
Charge Nurse - Newborn
January 2008:
Updated Dec 08 (Scavenging System)
Clinical Guidelines Back Newborn Services Home Page
Administration Equipment Required Process

Administration via Drager Babylog (in all modes) or SensorMedics

Equipment Required

Nitric Oxide Accessory Pack for Drager Babylog and Sensormedic You will also require

Assemble Circuit

Set Up for Nitric Oxide via Drager Ventilators using Printer NOx

NO-Drager.jpg (37189 bytes)
Click on the small image to see a larger image

NO via SensorMedics HFOV showing Connections Needed

NO SensorMedics HFV Connections needed.jpg (56240 bytes)
Click on the small image to see a larger image

Scavenging System for Drager

Click on the small image to see a larger image

The scavenger has no pressure gauge. Once scavenger circuit is connected, turn dial until suction noise is audible.

Scavenging System for SensorMedics

Click on the small image to see a larger image

 

  

Process

Step

Action

1

Ensure circuit is set up correctly as per appropriate diagram:
  • Sampling line attached to cylinder → nitric adaptor and placed in circuit as per diagram
  • Printer NOx plus sampling line attached to nitric adaptor.
  • Scavenging unit in place (see above)

2

Doctor/NS-ANP is present when NO flow is turned on:
  • On cylinder: either one knob or two knobs - middle knob first (different cylinder heads in service)
  • Flow meter: 0.2 litres (200mls) per min with a gas flow of 10 litres per minute gives around 17ppm (check approximate concentration using the nitric oxide calculator)

 NB:

The NO cylinder is checked for leaks using a "snoop liquid leak detection" (instructions on   bottle)
  • when regulator is attached to a new cylinder
  • if leak suspected
 