Calibration of the MAXO Oxygen Analyser

 

Reviewed by Nicola Svirskis
April 2010
Clinical Guidelines Back Newborn Services Home Page

Steps Action

1

Remove analyser from respiratory circuit for calibration. Occlude port in circuit to ensure continued flow to infant.

2

If the infant is receiving <60% oxygen then calibrate to 9room) air.

 

If the infant is receiving >60% oxygen the analyser needs to be calibrated to 100%oxygen.

  • plug the analyser sensor to the calibration jig (from the equipment room)

  • plug in a flow regulator to the oxygen port on the ponta and attach the calibration jig 

  • attach oxygen calibration jig to sensor

  • set flow rate at 2 L/min of oxygen

Click image to enlarge

3

Turn unit on (on/off key).
4 Press ‘lock’ key to unlock unit.
5 Press ‘cal’.
6 Analyser will calibrate automatically (adjust á or â keys if required, till calibration figure reached).
  • if calibrating to air set analyser to 21%
  • if calibrating to pure oxygen set analyser to 100%
7 When calibration complete, analyser unit will lock automatically.  Replace sensor into circuit.