| Steps
|
Action
|
|
1 |
Remove analyser from respiratory circuit for calibration. Occlude port in
circuit to ensure continued flow to infant. |
|
2 |
If the infant is receiving <60% oxygen then
calibrate to 9room) air.
If the infant is receiving >60% oxygen the
analyser needs to be calibrated to 100%oxygen.
-
plug the analyser
sensor to the calibration jig (from the equipment room)
-
plug in a flow
regulator to the oxygen port on the ponta and attach the calibration
jig
-
attach
oxygen calibration jig to sensor
-
set flow rate at 2
L/min of oxygen
 |
| Click image to enlarge |
|
|
3 |
Turn
unit on (on/off key). |
| 4 |
Press ‘lock’ key to unlock unit. |
| 5 |
Press
‘cal’. |
|
6 |
Analyser
will calibrate automatically (adjust á
or â
keys if required, till calibration figure reached).
- if calibrating to air set analyser to 21%
- if calibrating to pure oxygen set analyser to 100%
|
|
7 |
When calibration complete, analyser unit will lock automatically.
Replace sensor into circuit. |