Blue water goes up the narrow
water seal tube (e.g. to 10-15cm)
|
This is due to
a rise in the intrathoracic pressure. The can be caused by using pressure to remove air by syringe from the chest
(you should not need to have a syringe on the 3 way tap use a leur cap).
To correct
: Depress the manual high negativity vent located on the top of the
drain until the water seal column lowers to the desired level. Do not
lower water seal column when suction is not operating, or when on
gravity-drainage only. Temporarily connect to the wall suction first.
|