Intravascular Catheters
|
Authorised by: Charge Nurse - Newborn |
| May 2006 |
|
Steps |
Action |
|
1 |
Specific equipment required:
|
|
2 |
Observe baby carefully for any bruising on buttocks, toes and feet and document carefully. |
|
3 |
Ensure baby has been measured from umbilicus to shoulder to give approximate catheter placement. |
|
4 |
Secure baby’s limbs (see Restraint
guideline). After insertion ascertain from NS-ANP/Medical staff which catheter is a UAC and UVC and label clearly. |
|
5 |
A radiograph is used to confirm correct position of
UAC/UVC. UAC fluids and blood pressure measuring can be commenced before the radiograph is taken. |
|
6 |
Do not lie baby prone for at least 4 hours after insertion. Do not lie baby prone unless UAC/UVC lines are secured using the Gate Strapping method. |


|
Step |
Action |
|
1 |
To secure catheter wrap one suture thread around the catheter x 2 firmly and close to the base. Double knot and repeat. |
|
2 |
Take a 5cm length of elastoplast. Attach to the catheter – close to the base. Position the threads along the length of the catheter and wrap half the tape around and around. |
|
3 |
Fold back the thread lengths and continue wrapping the tape around catheter, catching the threads again. |
|
4 |
At the end, turn under the last bit of tape to make a tag. |

|
Steps |
Action |
|
1 |
Place duoderm base tapes lengthwise on the baby’s abdomen, one on either side of the umbilicus. Then apply sleek gates on top of duoderm. Size of gate should be relevant to size of infant. |
|
2 |
Until a radiograph for catheter position is performed, secure catheter with a temporary crossbar using leucoplast. |
|
3 |
After the radiograph, the catheter may need repositioning. Once this has been done restrap, using sleek as the crossbar to secure. |
|
N.B.: |
|
|
Steps |
Action |
|
1 |
A nurse is always present in the room. Alarm limits are set appropriately including mean blood pressure alarm. |
|
2 |
The catheter is secured as per baby’s weight criteria. Check security of gate at the commencement of each duty. Retape or apply a sleek gate immediately if concerned about dislodgement. |
|
3 |
Connections are checked at beginning of each shift to ensure they are secure and there are no kinks in the lines. |
|
4 |
Booties are not
worn. The toes, feet and legs are checked frequently for changes in colour
and circulation. Buttocks are checked for patchy discolouration and signs of blanching. Cyanosis of toes/feet/buttock discoloration are reported to doctor/NS-ANP immediately. |
|
5 |
Napkins must not be fastened in such a manner as to obscure view of the umbilical stump. |
|
6 |
Babies total fluids must not be infused via a UAC with a blood pressure transducer in situ as the flush device has a minimum flow rate. |
|
N.B.: |
|
|
Steps |
Action |
|
1 |
Cut stitches between catheter and knot of purse string. |
|
2 |
Withdraw catheter slowly until 3cm remains in UAC/UVC. |
|
3 |
Retape catheter at that position. |
|
4 |
Turn 3 way tap off to catheter. |
|
5 |
Turn off infusion pump. |
|
6 |
Leave for 30 minutes. |
|
7 |
Remove catheter completely. |
|
N.B.: |
Have sterile gauze handy to apply pressure if bleeding occurs. |
|
8 |
Do not lie baby prone for at least 4 hours after removal of UAC/UVC. Check frequently for ooze. |
|
9 |
Ensure napkin is secured below the umbilicus. |
|
Steps |
Action |
|
1 |
Ensure there is no blood present in the blood pressure line causing damping of the reading. |
|
2 |
Calibration checks are performed - at the beginning of each shift or if having problems with the readout of blood pressure. |
| Equipment required |
|
|
Steps |
Action |
|
1 |
Attach blood pressure transducer set to giving set. |
|
2 |
Plug cable from monitor into BP transducer set. A labelled pressure waveform is displayed when the cable is connected |
|
3 |
Prime giving set. |
|
4 |
Squeeze lever on flush device and prime BP set. |
|
5 |
Turn stopcock off to flushing line and syringe. |
|
6 |
Remove the 10ml syringe, draw up 10mls NaCl with Heparin from IV Bag, replace and prime flushing line to the end of transducer set. |
|
Note: |
Peripheral arterial line. Prime line as above, then remove syringe and attach luerlock bung. |
|
7 |
Turn stop cock off to the flushing line and syringe. Check that there are no air bubbles present in transducer line and flushing line. |
|
Note: |
Transducer must have a zero calibration performed prior to commencement of BP monitoring. A zero calibration is required once per shift. |

|
Steps |
Action |
|
1 |
Ensure transducer placed appropriately at level of mid-axilla of baby. |
|
2 |
Close the 3-way tap closest to the baby. |
|
3 |
Open the venting stopcock on the transducer to air. |
|
4 |
Press the "Zero all" button on the keypad |
|
5 |
Verify that zero reference has been established (watch the pressure parameter window for messages) |
|
6 |
Close the venting stopcock. |
|
7 |
Open the 3 way tap to baby. Within seconds, pressure numerics should be displayed in the pressure parameter window. |
Disconnect ECG Leads
↓
Discharge Baby from Monitor
↓
Re-admit baby to Monitor
(Monitor will set up automatically)
If you cannot see an index above, please
press the "Refresh" button of your browser.