Ministry of Health
NZ Government

©Copyright
Published:
29/11/2011
|
The "Panting" Alarm
|
Reviewed by Carl Kuschel and
Robyn Wilkinson |
April
2002 |
The Babylog 8000plus may
occasionally alarm with a mesage "Frequency High!". This will generally
alarm when the ventilator recognises respiratory rates above the predetermined
limit. The alarm may trigger because the baby truly is this tachypnoeic,
but it most commonly occurs because there is water in the ventilator tubing.
This gives the sensor the signal that the baby is breathing when he is not, and
the ventilator will auto-trigger.
The alarm limits for the "Panting"
alarm are generally set at 100 breaths per minute. DO NOT CHANGE THIS
SETTING WITHOUT FIRST FOLLOWING THE STEPS BELOW.
- Immediately respond by observing
the infant's respiratory rate in relation to the ventilator breaths being
provided.
- Is it synchronised?
- Note that on
Pressure Support Ventilation (PSV), the baby will get assisted breaths
at the rate the ventilator is autotriggering and will therefore appear
to be breathing quickly.
- Some babies on PSV will
breath rapidly when first placed on PSV. Generally they have also
been breathing quickly on SIMV. Give them some time to settle - if
they are still breathing quickly, you may need to adjust your settings
or find a different mode that suits.
- Is there water in the tubing?
- This is the most likely
reason for the rapid rate.
- The waves on the respiratory
function monitor will point towards this. There will be a
disturbed waveform on the pressure and flow waves.
 Pressure Wave form |
 Flow Wave form |
- Even a small amount of water can
occasionally be enough to autotrigger the ventilator.
- If there is no water in the
circuit and the alarm continues, seek advice.
- One option may be to
increase the trigger sensitivity
Other Ventilation Links
|